LYS-5
Lyse reagent for Spincell 5 Series
IVD
INTENDED USE
This product is designed to be used on a Spinreact 5-Part-Diff
haematology analyser along with Lyse and Sheath produced by
SPINREACT, for testing the cell content in human blood
FUNCTIONS
Lyse does not contain trinitride or cyanide. It has the functions of
lysing the red cells and releasing the hemoglobin to protect the RBC
count from being affected by RBC fragments. At the same time it
converts hemoglobin to steady colored hemo-compound for detecting
the concentration of hemoglobin. The Lyse preserves the integrity of
WBC nucleus allowing the counting and classification of WBCs
INGREDIENTS
Dodecyl trimethyl ammonium chloride 25~30 g/L
Myristyltrimethylammonium bromide 1,0~2,0g/L
Hydroxylamine hydrochloride 1,0~2,0 g/L
PRECAUTIONS
H290-May be corrosive to metals. EUH208-Contains Hydroxylamine
hydrochloride. May produce an allergic reaction.
Follow the precautionary statements given in MSDS and label of the
product.
STORAGE & STABILITY
This product will be stable for 24 months when stored unopened at
2℃~40℃ and away from sunlight. Once opened, this product should
be consumed within 60 days and stored below 35℃
BLOOD SAMPLES
1. The blood samples should be venous and capillary
specimens collected in EDTA-K2 anticoagulant. Coagulated,
hemolysis and lipoidaemia samples cannot be measured.
2. Blood samples should be stored at 2℃~8℃.
OPERATIONS
1. Place the Lyse container beneath the analyzer to avoid
direct sunlight. Keep away from cooling or heating air outlet.
2. Use it at 15℃~30℃.
3. Connect Lyse container with the analyzer according to
Operation Manual.
TEST RESULT
If the test result is out of the range of reference value, flagged results
will be marked with letter “L” or “H”. “L” means the result is below the
lower limit; “H” means the result is higher than the upper limit.
RESTRICTIONS, PRECAUTIONS & NOTICE
1. Use the reagent under the professional technician’s
guidance.
2. Make sure that the reagent matches with your analyzer. Any
mismatch will lead to misleading results.
3. Avoid contacting with skin and eyes. In case of contact,
rinse immediately with plenty of water and seek medical
advice.
4. Avoid inhaling the evaporated reagent.
5. Tight up the container’s cap to prevent from volatilization and
contamination.
6. Once opened, dispose the remains over 60 days.
7. If product has been frozen, let product warm to room temperature,
and then gently invert product to mix it completely. Verify
background results before use.