Sheath reagent for Spincell 5 Series
IVD
INTENDED USE
This product is designed to be used on a Spinreact 5-Part-Diff
haematology analyser along with Lyse and Sheath reagents produced
by SPINREACT, for testing the cell content in human blood.
FUNCTIONS
Sheath is a buffer solution that allows cells to flow uniformly and avoid
aggregation. Its osmotic concentration and ionic strength keeps the
integrity and the volume of WBCs. It lyses the RBCs to form ghost
cells which would not scatter the laser light to avoid affecting WBC
measurement. Sheath spherizes WBCs and optimizes their scattering
property. It also has humidifying effect to prevent accumulation of air
bubbles in WOC flow system.
INGREDIENTS
Trihydroxymethyl Aminomethane <1 g/L
Hydrochlocric Acid <0,2 g/L
Triton <0,1 g/L
2-Phenoxyethanol <10 g/L
STORAGE & STABILITY
This product will be stable for 24 months when stored unopened at
2℃~40℃ and away from sunlight. Once opened, this product should
be consumed within 60 days and stored below 35℃
BLOOD SAMPLES
1. The blood samples should be venous and capillary
specimens collected in EDTA-K2 anticoagulant. Coagulated,
hemolysis and lipoidaemia samples cannot be measured.
2. Blood samples should be stored at 2℃~8℃.
OPERATIONS
1. Place the Sheath container beneath the analyzer to avoid
direct sunlight. Keep away from cooling or heating air outlet.
2. Use it at 15℃~30℃.
3. Connect Sheath container with the analyzer according to
Operation Manual.
TEST RESULT
If the test result is out of the range of reference value, flagged results
will be marked with letter “L” or “H”. “L” means the result is below the
lower limit; “H” means the result is higher than the upper limit.
RESTRICTIONS, PRECAUTIONS & NOTICE
1. Use the reagent under the professional technician’s
guidance.
2. Make sure that the reagent matches with your analyzer. Any
mismatch will lead to misleading results.
3. Avoid contacting with skin and eyes. In case of contact,
rinse immediately with plenty of water and seek medical
advice.
4. Avoid inhaling the evaporated reagent.
5. Tight up the container’s cap to prevent from volatilization and
contamination.
6. Once opened, dispose the remains over 60 days.
7. If product has been frozen, let it warm to room temperature, and
then gently invert the product to mix it completely. Verify the
background results before use.