PROCLEAN
Probe Cleaner for Spincell 3 and Spincell 5
Series
IVD
INTENDED USE
Probe Cleaner is appropriate for SPINREACT automated hematology
analyzers.
FUNCTIONS
Probe Cleaner is a reagent used to clean the instrument’s probe or
tubes. It can also be used for daily maintenance. The Probe Cleaner
does not impact the instrument’s use life, and will not stain or corrupt
the tubes.
INGREDIENTS
Chloride, demineralized water, hypochlorite.
STORAGE & STABILITY
This product will be stable for 12 months when stored unopened at
room temperature. Once opened, this product should be consumed
within 60 days.
OPERATIONAL PROCEDURE
When serious clogging occurs or instrument has not been properly
maintained for a time, the use of the probe cleaner is indicated.
1. Place a clean container with 2ml of probe cleaner reagent
under the sample probe.
2. Select “Clean” in the operator interface (follow the operation
manual). The instrument will aspirate probe cleaner
automatically.
RESTRICTIONS, PRECAUTIONS & NOTICE
1. Use the reagent under professional technician’s guidance.
2. Avoid contacting with skin and eyes. In case of contact,
rinse immediately with plenty of water and seek medical
advice.
3. Avoid inhaling the evaporated reagent.
4. Tight up the container’s cap to prevent volatilization and
contamination.
5. Once opened, dispose the remains over 60 days.
6. Avoid product freezing